Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00717-1
Product Name/Description Leica ASP200S & Leica ASP300S. In vitro diagnostic medical device (IVD)
(Histology sample processing instruments)

Leica ASP200S
Product Code: 14048043625
Serial Numbers: 5329, 5448, 5528, 5707

Leica ASP300S
Product Code: 14047643514
Serial Numbers: 5361, 5639, 5721, 5872, 6057

ARTG Number: 189287
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/08/2015
Responsible Entity Leica Microsystems Pty Ltd
Reason/Issue Leica Microsystems recently became aware that affected devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure. The missing alarm signal could lead to suboptimal quality tissue processing. In some cases, a re-biopsy of the patient could be necessary in order to obtain another viable tissue sample.
Recall Action Recall for Product Correction
Recall Action Instructions A Leica Service Engineer will correct the internal wiring and check the remote alarm functionality.

This action has been closed out on the 17/10/2018.