Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00717-1 |
Product Name/Description |
Leica ASP200S & Leica ASP300S. In vitro diagnostic medical device (IVD) (Histology sample processing instruments)
Leica ASP200S Product Code: 14048043625 Serial Numbers: 5329, 5448, 5528, 5707
Leica ASP300S Product Code: 14047643514 Serial Numbers: 5361, 5639, 5721, 5872, 6057
ARTG Number: 189287 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
5/08/2015 |
Responsible Entity |
|
Reason/Issue |
Leica Microsystems recently became aware that affected devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure. The missing alarm signal could lead to suboptimal quality tissue processing. In some cases, a re-biopsy of the patient could be necessary in order to obtain another viable tissue sample. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
A Leica Service Engineer will correct the internal wiring and check the remote alarm functionality.
This action has been closed out on the 17/10/2018. |