Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00715-1
Product Name/Description Volcano IVUS Ultrasound Imaging System

Volcano s5 and CORE Series

Software versions 3.3, 3.4 (with and without iFR Scout technology) and v3.4 software kits
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/08/2015
Responsible Entity Medical Vision Aust Cardiology & Thoracic Pty Ltd
Reason/Issue Volcano Corp. has discovered a very specific workflow scenario where the Aortic pressure (Pa) offset used in the FFR/iFR modalities will be lost or reset to 0 mmHg due to a software error. This issue can lead to an inaccurately calculated FFR or iFR value. Should this occur, a clinician may rely on an inaccurate value in diagnosing a patient's condition.

If the Pa was not zeroed through the Volcano system as part of the workflow, this software issue has no impact on the FFR / iFR values or on the patient.
Recall Action Recall for Product Correction
Recall Action Instructions Medical Vision Australia is advising their customers that if zeroing Pa is warranted, as described in the FFR Option Operator's Manual, performance of this product cannot be assured until the software update is provided. A Medical Vision Australia Representative will contact affected hospitals to schedule a correction to the software in the next few weeks. This action has been closed-out on 24/08/2016.
Contact Information 1300 661 559 - Medical Vision Australia