Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00682-1 |
Product Name/Description |
N Latex ß2–Microglobulin. An in vitro diagnostic medical device (IVD) (Quality control reagent)
Catalogue Number: OQWU155 Kit Number: 44696, 44800, and 44834 Catalogue Number: OQWU215 Kit Number: 44697
ARTG Number: 178153 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
23/07/2015 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has confirmed complaints about Calibration/QC validation issues as well as 'Abnormal Assay Error' flags when using the affected lots. Affected reagent vials and reagent flexes show partial flocculation and share the same production bulk. |
Recall Action |
Recall |
Recall Action Instructions |
Siemens is requesting their customers to discontinue use and discard the affected lots. This action has been closed-out on 06/07/2017. |
Contact Information |
1800 310 300 - Technical Support Center |