Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00682-1
Product Name/Description N Latex ß2–Microglobulin. An in vitro diagnostic medical device (IVD)
(Quality control reagent)

Catalogue Number: OQWU155 Kit Number: 44696, 44800, and 44834
Catalogue Number: OQWU215 Kit Number: 44697

ARTG Number: 178153
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 23/07/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has confirmed complaints about Calibration/QC validation issues as well as 'Abnormal Assay Error' flags when using the affected lots. Affected reagent vials and reagent flexes show partial flocculation and share the same production bulk.
Recall Action Recall
Recall Action Instructions Siemens is requesting their customers to discontinue use and discard the affected lots. This action has been closed-out on 06/07/2017.
Contact Information 1800 310 300 - Technical Support Center