Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00677-1
Product Name/Description Artis systems used in conjunction with Artis table

Catalogue and Serial number – Multiple serial number ranges of Artis tables with the following Siemens Material Numbers:

5767004
5767012
5767020
5767053
10759275
10759276
10759277
10759278

ARTG number – 102173 and 102177
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/07/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue For Artis systems used in conjunction with Artis table;-

In the event the tabletop is approximately fully extended at the head end and there is a large amount of liquid on the table surface, the liquid may make its way from the foot end of the tabletop onto the top cover of the table column where it can seep through a gap. This liquid may then cause interference with electronic components and table movements may no longer be possible.
Recall Action Recall for Product Correction
Recall Action Instructions The gap in the top of the table cover will be sealed to ensure protection against the entry of liquids.
Contact Information 1800 310 300 - Technical Support Centre