Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00677-1 |
Product Name/Description |
Artis systems used in conjunction with Artis table
Catalogue and Serial number – Multiple serial number ranges of Artis tables with the following Siemens Material Numbers:
5767004 5767012 5767020 5767053 10759275 10759276 10759277 10759278
ARTG number – 102173 and 102177 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
28/07/2015 |
Responsible Entity |
|
Reason/Issue |
For Artis systems used in conjunction with Artis table;-
In the event the tabletop is approximately fully extended at the head end and there is a large amount of liquid on the table surface, the liquid may make its way from the foot end of the tabletop onto the top cover of the table column where it can seep through a gap. This liquid may then cause interference with electronic components and table movements may no longer be possible. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
The gap in the top of the table cover will be sealed to ensure protection against the entry of liquids. |
Contact Information |
1800 310 300 - Technical Support Centre |