Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00671-1 |
Product Name/Description |
iChemVELOCITY Automated Urine Chemistry System using iChemVELOCITY Urine Chemistry Strips
Strip part numbers 800-7204 and 800-7204-001 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
21/07/2015 |
Responsible Entity |
|
Reason/Issue |
The iChemVELOCITY systems do not maintain the on-board strip stability claim of 10 days at 18-28 degrees Celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labelling. The on-board strip stability is affected by temperature and humidity in your laboratory. Test results may be affected as indicated below:
False low or false negative results: Ketone, Blood False high or false positive results: Bilirubin, Urobilinogen, Glucose, Nitrite False high results: Leukocytes
Ascorbic Acid, Protein and pH are not affected. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter is providing an updated stability guide and additional handling instructions to ensure the stability of the test strips. The manufacturer, Iris International, has determined that the Strip Provider Module is not maintaining the humidity levels associated with the original on-board strip stability claim and is urgently developing a corrective action. This action has been closed-out on 06/09/2016. |
Contact Information |
1800 060 881 - Customer Support Centre |