Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00671-1
Product Name/Description iChemVELOCITY Automated Urine Chemistry System using iChemVELOCITY Urine Chemistry Strips

Strip part numbers 800-7204 and 800-7204-001
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 21/07/2015
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue The iChemVELOCITY systems do not maintain the on-board strip stability claim of 10 days at 18-28 degrees Celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labelling. The on-board strip stability is affected by temperature and humidity in your laboratory. Test results may be affected as indicated below:

False low or false negative results: Ketone, Blood
False high or false positive results: Bilirubin, Urobilinogen, Glucose, Nitrite
False high results: Leukocytes

Ascorbic Acid, Protein and pH are not affected.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is providing an updated stability guide and additional handling instructions to ensure the stability of the test strips. The manufacturer, Iris International, has determined that the Strip Provider Module is not maintaining the humidity levels associated with the original on-board strip stability claim and is urgently developing a corrective action. This action has been closed-out on 06/09/2016.
Contact Information 1800 060 881 - Customer Support Centre