Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00626-1
Product Name/Description Artis Q and Q.zen Floorstand and ArtisQ and Q.zen Biplane Floorstand systems with SW version VD10E

Catalogue/Lot number – 10848280, 10848355, 10848282 with SW version VD10E

ARTG Number: 102177
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/07/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue If a position sensor fault in the swivel base axis is not detected by the system software, the C-arm system may move faster than normal when movement is enabled by the operator. The system will not move of its own accord, movement must still be initiated by the operator, however the system can unexpectedly exceed its usual speed, potentially resulting in a collision. There have been no complaints from Australian customers regarding this issue.
Recall Action Recall for Product Correction
Recall Action Instructions This issue will be remedied with software update AX037 /15/S (Patch 2). Following the installation of this update, the speed of the C-arm will be limited to a pre-determined level. This action has been closed-out on 12/08/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre