Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00599-1
Product Name/Description ACUSON S1000, ACUSON S2000, or ACUSON S3000 Ultrasound Systems with software version VC30A, VC30B, VC30C, or VC31A

ARTG Number: 137563
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/07/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has recently discovered a potential miscalculation when using the Doppler manual trace measurement tool in a full screen display format. The manual trace tool may incorrectly calculate the Time Average Velocity (TAV) and subsequently incorrectly calculate the Pulsatility Index (PI). The potential risk is a misdiagnosis however, the probability is low.
Recall Action Recall for Product Correction
Recall Action Instructions This issue will be resolved in a future software release. In the interim to avoid this issue, users are advised not to perform measurements utilising the manual Doppler trace tool in a full screen display format. This action has been closed-out on 15/02/2017.