Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00579-1 |
Product Name/Description |
Parechovirus r-gene. An in vitro diagnostic medical device (IVD)
Catalogue Number: 71-020
Serial Number: 1002978600
Supplied as Research Use Only
ARTG Number: 211361 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
29/06/2015 |
Responsible Entity |
|
Reason/Issue |
An internal investigation identified a shift of the Ct values of lnternal Control (IC1 W0, Internal Control with water) inside and outside the limit of specifications.
The investigations performed by R&D with ref. 71-020 lot 1002978600 showed that Ct values results on the extraction control (IC1W0) with Ct value close to the specification of validity (32 Ct) or just above this specification. The Ct shift values may be observed on a positive sample without impacting the limit of detection. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Biomerieux is requesting their customers to discontinue using and discard any remaining inventory of the Parechovirus r-gene ref. 71-020 (WW) lot 1002978600. This action has been closed-out on 17/08/2016. |