Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00578-1
Product Name/Description Milestone Wand STA Hand Piece model STA-5050-305

Batch or Lot number:140103

Expiry date: January 2019

ARTG Number: 95633
Recall Action Level Retail
Recall Action Classification Class III
Recall Action Commencement Date 8/07/2015
Responsible Entity Henry Schein Halas
Reason/Issue Wand hand piece model WA-2050-305 was packaged (pouched) and boxed as a Wand STA hand piece model STA-5050-305.

Although the function of the hand pieces are identical, these hand pieces are used in two different instruments and they will only attach into the correct instrument.
Recall Action Recall
Recall Action Instructions Milestone is requesting the customers to inspect the inventory of the Wand Handpieces by looking at the cartridge holder locking tab without opening the sterile pouch. If any mislabelled handpieces are identified, Milestone is requesting the customers to contact Milestone Customer Service to obtain return authorisation and / or replacement of the handpieces. Milestone is also instructing the customers that they can also return the entire quantity of the affected STA Handpieces. This action has been closed-out on 31/08/2016.
Contact Information 1300 65 88 22 - Customer Care