Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00488-1
Product Name/Description Access 2 Immunoassay Systems
(Includes the Access 2 Immunoassay System and UniCel DxC 600i Synchron Access Clinical System)

An in vitro diagnostic device (IVD)

All Serial Numbers

All Software Versions
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/06/2015
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Beckman Coulter updated the Instructions for Use (IFU) to include the recommended temperature restrictions to address the impact of room temperature changes on certain assays run on Access 2 Immunoassay Systems. The new updated operation range temperatures are between 18 to 28 degrees Celsius. An updated version of the IFU is available on the Beckman Coulter website ( http://www.beckmancoulter.com)
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is notifying their customers that the updated version of the Access 2 Immunoassay Systems IFU is available on the Beckman Coulter website ( http://www.beckmancoulter.com) This action has been closed-out on 19/08/2016.