Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00461-1
Product Name/Description Verathon GlideScope GVL and AVL Reusable Video Laryngoscope Blades

GVL 3
Part Number( PN): 0574-0007
Serial Number Range(SN) : MD131860 - MD151622

GVL 4
PN: 0574-0001
SN: LG131995 - LG151773

GVL 5
PN: 0574-0030
SN: XL131578 - XL151524

AVL 2
0574-0118
AC131852 - AC151510

AVL 3
0574-0115
AD131589 - AD151506

AVL 4
0574-0116
AE131587 - AE151558

AVL 5
0574-0117
AF131504 - AF151500

ARTG Number: 235136
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/06/2015
Responsible Entity Verathon Medical Australia Pty Ltd
Reason/Issue Verathon, manufacturer of Glidescope video laryngoscopes, is revising the Operations and Maintenance Manuals (OMM) for certain Glidescope GVL and AVL Reusable Laryngoscope Blades. The new OMMs specifically state compatibility limitations of the reusable video laryngoscope validated through simulated use cycles, using certain cleaning and disinfection agents. Also, for the purpose of cleaning/disinfecting reusable blades, the following chemicals have been removed from the OMMs:

• Bleach (sodium hypochlorite)
• Isopropyl alcohol solution
• Liquid hydrogen peroxide (7.5%)

There is no change to the safety information provided in the OMM regarding inspection of the laryngoscope blade before and after every use to ensure patient safety.
Recall Action Recall for Product Correction
Recall Action Instructions Verathon is informing users regarding updates to the following Operations and Maintenance Manuals:

• GVL Operations and Maintenance Manual: 0900-1204-08-60
• AVL Operations and Maintenance Manual: 0900-4200-02-60

The updated Operations and Maintenance Manuals can be found at www.verathon.com. This action has been closed-out on 11/08/2016.
Contact Information 1800 613 603 Verathon Medical