Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00430-1
Product Name/Description Dimension Vista 500 and 1500 Intelligent Lab System. An in vitro diagnostic medical device (IVD)

Software version: 3.6.1, 3.6.1_MU3p and 3.6.1SP1

ARTG Number: 174699
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 20/05/2015
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Issue 1:
Samples may stop processing without notification on the Dimension Vista 500 or Dimension Vista 1500. This issue is due to a failure to exchange aliquot plates, causing the lane to become unavailable.

Issue 2:
Complaints were received on the Dimension Vista 1500 for a series of unexpected results, some of which were not flagged with errors or results that were flagged with errors. It has been determined that under a rare set of conditions, a timing issue may cause a Reagent Server to temporarily lose synchronisation during the automatic removal of reagent cartridges from Reagent Server 2 to Waste A Container. In the extremely unlikely circumstance that this issue occurs, a flag will be generated for the majority of assays. However, there is the potential to produce unflagged erroneous but believable results for some assays.
Recall Action Recall for Product Correction
Recall Action Instructions Users are provided with temporary workarounds in the customer letter. Siemens will be providing corrections for the issues in a future Vista software version. Review of previously generated results due to this issue is at the discretion of laboratory. This action has been closed-out on 26/08/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre