Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00371-1
Product Name/Description Transwarmer Infant Transport Mattress

Catalogue number: 20421

All batches

Order codes: BPOD2-TRANS06 / BPOD2-NEON03

ARTG number: 221364
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/04/2015
Responsible Entity Ferno Australia Pty Ltd
Reason/Issue CooperSurgical has identified that there may be product with customers that have been in inventory for more than five years. Some users of the TransWarmer Infant Transport Mattress may have removed individual units from the original packaging and have been using the device past the labelled and recommended expiration date.

Serious burns have been reported in the past and could occur if the products are not stored under the labelled conditions or used beyond the recommended shelf life. CooperSurgical has received three complaints from the field indicating that the product may activate at higher than labelled temperatures.
Recall Action Recall
Recall Action Instructions CooperSurgical is requesting all users to review their inventory and discard or return any product that does not display a product LOT code or expiration date on each individual unit. The brands affected include CooperSurgical and NovaPlus labelled TransWarmer Infant Transport Mattress. The LOT code and expiration date can be located on the top left corner of the individual pouches.

All new units will have the lot and expiry date on the individual unit packaging. This action has been closed-out on 17/05/2017.
Contact Information 1800 804 647 - Ferno Australia