Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00370-1
Product Name/Description HeartWare HVAD System

Product Codes: 1100, 1101, 1102, 1104, 1205
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 7/05/2015
Responsible Entity Heartware Pty Ltd
Reason/Issue HeartWare is providing further information to users regarding:
1. Worn Alignment Guides,
2. Internal "Double Disconnect Alarm" Battery,
3. Power Management Software Upgrade,
4. Driveline Outer Sheath Discoloration & Cracking, and
5. Snagging or Pulling of the Driveline.
Recall Action Recall for Product Correction
Recall Action Instructions HeartWare is reiterating warnings and precautions stated in the Instructions for Use and Operators Manual to ensure the safe use of the devices. HeartWare is also notifying health care professionals of an additional warning regarding discolouration and cracking of the drive line due to exposure to UV light.

Further information can be found ont he TAG website at https://www.tga.gov.au/alert/heartware-ventricular-assist-system

This action has been closed out on the 26/10/2018.
Contact Information 02 8078 6164 - HeartWare Pty Ltd