Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00368-1
Product Name/Description Factor IX coagulation studies on the BCS / BCS XP Automated Blood Coagulation Analyser (Actin, Actin FSL and Pathromtin SL)

Siemens Material Numbers (SMN):
BCS: 10454742, 10454729; BCS XP: 10461894, 10470625

ARTG Number: 178116

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/04/2015
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens has detected a situation with BCS / BCS XP analysers regarding Factor IX assay determinations for Haemophilia B patients. Internal investigations showed discrepancies for Siemens BCS / BCS XP within dilution studies. Siemens has confirmed that samples at the lower measurement range were found elevated on the BCS/BCS XP system.
Patient results in the lower measuring range may be misclassified within a higher Factor IX level. Patients that have been evaluated for hereditary factor IX deficiency disease for the first time may have to be re-evaluated.

A falsely elevated factor IX level may lead to a delay in diagnosis of a bleeding disorder related to factor IX deficiency. This may lead to additional diagnostic testing and delayed prophylactic therapy.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens has provided a workaround for users along with a technical support contact number, whilst a permanent fix is being developed. Customers are advised to follow their laboratory protocol regarding the need for review of previously reported patient results.
This action has been closed-out on 01/03/2018.
Contact Information 1800 310 300 - Siemens Technical Support Center