Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00360-1
Product Name/Description Thermablate Endometrial Ablation System (EAS)

Thermablate Treatment Control Unit Kit
Product Code: I-22001

Thermablate Disposable Cartridge
Product Code: 1-21004

All Lot Numbers

ARTG Number: 186661
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/04/2015
Responsible Entity Gytech Pty Ltd
Reason/Issue The Instructions For Use (IFU) of the Thermablate Endometrial Ablation System (EAS) is updated due to four case reports of uterine perforation. There is an increased risk of patient injury (including damage to non-targeted tissue) if both Thermablate EAS procedure and hysteroscopic tubal occlusion/sterilisation are performed on the same day.
Recall Action Recall for Product Correction
Recall Action Instructions Additional warning is being added to the IFU as follows :
"Do not perform same day Thermablate EAS procedure and hysteroscopic tubal occlusion/sterilization. Thermablate EAS procedure can be safely and effectively performed with nickel titanium inserts in place, however the procedure should only be performed after the 3 month tubal occlusion confirmation test."

Gytech Pty Ltd is informing the customers of the additional warning as updated in the Instructions For Use (IFU). The IFU document (Ref. No: LS2101, Rev C) currently supplied with the Disposable Cartridge already includes these changes. This action has been closed-out on 5/08/2016.
Contact Information 03 9822 5911 - Gytech Pty Ltd