TGA Recall Reference RC-2015-RN-00331-1
Product Name/Description Dako Beta-2-Microglobulin PET Kits. An in vitro diagnostic medical device (IVD)

Product code: K0052
Lot Number: 20012249

ARTG number: 200116
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/04/2015
Responsible Entity Elitechgroup Australia PTY LTD
Reason/Issue There is a stability issue with a single component of the kit, Vial 1 Beta-2-microglobulin Immunoparticles. Investigations have shown that a change in the pH value in the Beta-2-microglobulin Immunoparticles (Vial 1) has occurred resulting in low or no signal for the 2 lowest standards in the calibration curve. The identified change in the pH value may affect the accuracy of Beta-2-microglobulin values/results obtained, without any warnings from instruments. The Beta-2-microglobulin values may be measured up to 20% higher for samples in the clinically elevated ranges.
Recall Action Recall
Recall Action Instructions ELITech Group is advising users to cease use of the affected lot and discard any remaining inventory. This action has been closed-out on 18/07/2016.
Contact Information 1800 815 098 - ELITechGroup Australia