Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00302-1 |
Product Name/Description |
Nobel Replace Conical Connection PMC RP 5.0x13mm ( Fully Imbedded Dental Implant – Screw endosteal) Batch number: 13010296 Article number: 37298 ARTG Number: 202090 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
2/04/2015 |
Responsible Entity |
|
Reason/Issue |
Nobel Biocare Quality Management Systems observed an issue with a material used in the packaging of the implants. The packaging component (cap) showed excessive oil which was caused by an issue in production of the raw material (pellets) for manufacturing the caps. Nobel Biocare's internal analysis has confirmed that this material is not according to the standards. Based on the testing performed so far the likelihood that the products cause adverse health consequences is currently considered to be low. The implant batches still at the manufacturing site were restricted for further distribution and placed in quarantine. |
Recall Action |
Recall |
Recall Action Instructions |
Nobel Biocare requested the customer to return the affected product with batch number 13010296. Nobel Biocare is providing a replacement product. This action was underatken prior to notifying TGA. |