TGA Recall Reference RC-2015-RN-00201-1
Product Name/Description IMMULITE 2000/IMMULITE 2000 XPi water and liquid waste bottles

Siemens Material Numbers: 10291586 and 10291587

ARTG number: 174399
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/03/2015
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics is conducting a correction for the IMMULITE 2000 / IMMULITE 2000 XPi Immunoassay Systems water and liquid waste bottles received from May 2013. Siemens has identified that the water and liquid waste bottle assemblies were manufactured with a quality issue. The opening of both bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.

This could result in the aspiration straw dislodging from the bottle causing the instrument to aspirate air into the fluidic line potentially resulting in discordant results.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing work around instructions to assist end users to mitigate risk and identify faulty bottle openings. If a faulty bottle is identified end users are requested to contact Siemens for a replacement.

This action has been closed out on 26/09/2018
Contact Information 1800 310 300 - Siemens Technical Support Centre