Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00183-1
Product Name/Description Hemodialysis Monitor

Catalogue number: HD03

Lot numbers: A81225, A91394 & B21812

ARTG number: 187558
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/03/2015
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue This recall is being initiated because of customer complaints indicating the inability to obtain measurements due to the device sometimes seeing no flow or irrelevant flow condition when used with patients during dialysis procedures. No reports of any injuries have been reported nor are expected due to the nature of this issue.

The problem is readily identified to the user by a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).
Recall Action Recall
Recall Action Instructions Emergo is advising their customers to discontinue the use of the monitor and contact Transonic customer service to arrange for replacement (repair/exchange) immediately. This action has been closed-out on 18/07/2016.