Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00183-1 |
Product Name/Description |
Hemodialysis Monitor
Catalogue number: HD03
Lot numbers: A81225, A91394 & B21812
ARTG number: 187558 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
19/03/2015 |
Responsible Entity |
|
Reason/Issue |
This recall is being initiated because of customer complaints indicating the inability to obtain measurements due to the device sometimes seeing no flow or irrelevant flow condition when used with patients during dialysis procedures. No reports of any injuries have been reported nor are expected due to the nature of this issue.
The problem is readily identified to the user by a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output). |
Recall Action |
Recall |
Recall Action Instructions |
Emergo is advising their customers to discontinue the use of the monitor and contact Transonic customer service to arrange for replacement (repair/exchange) immediately. This action has been closed-out on 18/07/2016. |
Contact Information |
02 9006 1662 - Emergo Australia |