Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00161-1
Product Name/Description BD Q-Syte Luer Access Split Septum Device
(Connector designed to allow access to injection sites for delivery of medication)

Catalogue Number: 385100

Lot Number: 4128925

ARTG Number: 96767
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/02/2015
Responsible Entity Becton Dickinson Pty Ltd
Reason/Issue Some units have an off-centred slit in the silicon septum resulting in a thinner column wall. When actuated, units with a thinner column wall may tear leading to leakage of infusate/medication or blood from the side vent opening. In addition to leakage, the manufacturer has identified that there is a small
potential risk for air to enter the bloodstream which could result in air embolism.
Recall Action Recall
Recall Action Instructions Customers are advised to discontinue use of and immediately quarantine any of the affected stock. This action has been closed-out on 09/08/2016.
Contact Information 1800 656 100 - Becton Dickinson Customer Service Department