Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00157-1
Product Name/Description ONLINE DAT Cannabinoids II / 100 tests / 200 tests

Lot numbers: 698113, 604195, 604194, 696131, 698110 & 696133

ARTG number: 174911
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 6/03/2015
Responsible Entity Roche Diagnostics Australia Pty Limited
Reason/Issue Internal investigations by Roche Diagnostics verified that the within run precision (CV) for the positive/negative control with different reagent lots is approximately 10%. The specification is lower than 8%. Patient sample results which are very close to the cutoff might recover below the cutoff. The affected reagent lots expire in May 2015.

Newer lots are not affected by the precision issue and can be used as a replacement.
Recall Action Recall
Recall Action Instructions Roche is requesting their customers to discard any remaining stock of THC II from reagent lots 69811301 and 60419501. This action has been closed-out on 02/03/2016.
Contact Information 1800 645 619 - Roche National Support Centre