Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00146-1
Product Name/Description GemLock Long Hex Driver

Catalogue Number: TSVKIT

Lot Number on Label: 62702589*
Etched Lot Number of RHL2.5: 62699693
* Not all drivers used to assemble this kit lot are affected, as multiple lots of the driver were used to manufacture the kits. Drivers etched with lot 62632712, which function appropriately.

Lot Number on Label: 62743075
Etched Lot Number of RHL2.5: 62735177

ARTG number: 126570
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/02/2015
Responsible Entity Zimmer Pty Ltd
Reason/Issue Zimmer Dental identified that certain lots of the GemLock Long Hex Driver, Catalogue RHL2.5, may not fit into the Fixture Mount Transfer (FMT) or the internal hex of the implant. The Affected Product was also distributed within the Tapered SwissPlus & SwissPlus Implant Systems Kit, Surgical, Complete, catalogue OPCST as well as the Tapered Screw-Vent Implant System Surgical Kit, Complete, catalogue TSVKIT.
Recall Action Recall
Recall Action Instructions Zimmer is requesting their customers to quarantine the affected units. A Zimmer distributor will arrange for the removal of affected products. This action has been closed-out on 4/08/2016.
Contact Information 02 9483 5400 - Zimmer Pty Ltd