Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00126-1
Product Name/Description Philips Anaesthesia Machines Siesta i Breasy, Siesta i Whispa, Dameca MRI 508, Siesta i TS & IntelliSave AX700

Siesta i Breasy
Product Numbers: 866204/10652-00

Siesta i Whispa
Product Numbers: 866202/10651-00

Dameca MRI 508
Product Numbers: 866203/10651MRI-00

Siesta i TS
Product Numbers: 866163/10653-00

IntelliSave AX700
Product Numbers: 866205/10623-00

Multiple Serial Numbers affected

ARTG Number: 176848
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 24/02/2015
Responsible Entity Proact Medical Systems (Australia) Pty Ltd
Reason/Issue The spring-function of the control dial may not work, which may cause the control dial to get stuck when pushed, thus preventing its normal and expected function. It can be brought up to working condition again by simply pulling it out. The control dial is part of the user interface and serves an important role in setting and controlling the anaesthesia machine functions as it is used to adjust and confirm settings made by the user.

For Siesta i Breasy, Siesta i Whispa and MRI 508 freshgas regulation and ventilator mode change (Manual or Ventilator) can still be done as this is mechanical and/or soft key controlled and doesn’t need confirmation by the control dial.

For Siesta i TS and IntelliSave AX700 it is not possible to regulate freshgas delivery or change ventilator mode as this needs confirmation by the control dial.

For all machine types ventilation will continue at the previously set parameters
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is requesting all its customers to check their machines for the faulty control dial and is replacing control dials on devices exhibiting the issue. End users are provided work around instructions to assist in the continued use of affected units. The anesthesia machine can continue to be used as the control dial can be pulled out, making the control dial work. This action has been closed-out on 08/08/2016.
Contact Information 02 9975 5755 - Device Technologies Australia