Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00108-1
Product Name/Description Amulet (FDR MS-1000), Amulet f (FDR MS-2500) & Amulet Innovality (FDR MS-3500) with Biopsy Unit

Serial Numbers: 26340096, 37220054 & 36920025

ARTG Number: 159785
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/02/2015
Responsible Entity Fujifilm Australia Pty Ltd
Reason/Issue During a biopsy study using the FUJIFILM Digital Mammography System, it has been found that the User Manual lacks a description for handling some errors. If users do not address the recovery procedure properly, there may be a possibility that the examination is suspended leading to patient re-examination.
Recall Action Recall for Product Correction
Recall Action Instructions FujiFilm has developed a simplified manual describing the recovery procedure which will be circulated to customers. This action has been closed-out o 08/08/2016.
Contact Information 02 9466 2600 - FUJIFILM Australia Pty Ltd