Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00096-1 |
Product Name/Description |
VITEK 2 System version 7.01. An in vitro diagnostic medical device (IVD).
Bacteria identification and sensitivity system |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
4/02/2015 |
Responsible Entity |
|
Reason/Issue |
bioMeriuex has identified that the inoculum density range for the Neisseria Haemophilus (NH) Identification (ID) Test Card in the VITEK 2 7.01 Product Information document (Ref: 514740-1EN1) is incorrect. In Section 6 - Procedure For NH Card Specimen Preparation (step 3), the documented DensiCHEK Plus inoculum density range is 1.80 - 2.20. The correct density range is 2.70 - 3.30.
The incorrect information is located on the DVD software for the VITEK 2 7.01 system. All other product documentation has the correct inoculum density. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
bioMerieux is advising end users of the incorrect inoculum density range and requesting that the recall letter be kept with the instrument documentation. bioMerieux will update the VITEK 2 7.01 Product Information to include the correct DensiCHEK inoculation concentration. This action has been closed-out on 09/08/2016. |
Contact Information |
1800 333 421 - bioMérieux Australia - Technical Assistance Department |