Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00081-1
Product Name/Description CADD Medication Cassette Reservoir, 50ml

Order Code: 21-7001-24
Batch Numbers: 14X297 and 14X323

Order Code: 21-7301-24
Batch Numbers: 14X324

ARTG Number: 145297
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/01/2015
Responsible Entity Smiths Medical Australasia Pty Ltd
Reason/Issue Smiths Medical has received a small number of complaints of leakage associated with the above-referenced Cassettes. A small amount of leaking has been found to occur at the welding of the tube to the internal reservoir bag on some Cassettes. If the leakage is noticed during Cassette filling, it could result in a delay of therapy while an alternative Cassette is obtained. If the leakage is noticed during fluid delivery with the patient, it could result in an interruption in therapy while the Cassette is replaced.
Recall Action Recall
Recall Action Instructions End users are requested to inspect their inventory and quarantine the affected Cassettes. A customer service representative will arrange for exchange of any unused affected Cassettes for credit or replacement. This action has been closed-out on 17/08/2016.
Contact Information 02 9634 9200 - Smith's Medical Customer Service