Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00062-1 |
Product Name/Description |
ADVIA Chemistry Systems Drug Calibrator I for Phenytoin 2 and Phenobarbital 2 Assays. An in vitro Diagnostic Medical Device.
Lot Number: 3JD018
Expiration Date: April 2015
Siemens Material Number: 10376770
ARTG Number: 175699 |
Recall Action Level |
Retail |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
28/01/2015 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has determined that incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot 3JD018, for the ADVIA Chemistry Systems Phenytoin 2 (PHNY_2) and Phenobarbital 2 (PHNB_2) assays. This issue was determined to be the result of calibrator value miscalculation. Calibrator lot 3JD018 has now been value assigned to align with the correct calibrator values.
This poses a risk of an incorrect value of -10% being assigned for Phenytoin testing and 3% for Phenobarbital.
Siemens considers these differences are negligible and would not be considered substantially relevant to impact clinical care in practice. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Calibrator value reassignments are being provided in the Customer Letter for users to correct the problem. This action has been closed-out on 14/02/2017. |
Contact Information |
1800 310 300 - Siemens Technical Support Center |