TGA Recall Reference RC-2015-RN-00062-1
Product Name/Description ADVIA Chemistry Systems Drug Calibrator I for Phenytoin 2 and Phenobarbital 2 Assays. An in vitro Diagnostic Medical Device.

Lot Number: 3JD018

Expiration Date: April 2015

Siemens Material Number: 10376770

ARTG Number: 175699
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 28/01/2015
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics has determined that incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot 3JD018, for the ADVIA Chemistry Systems Phenytoin 2 (PHNY_2) and Phenobarbital 2 (PHNB_2) assays. This issue was determined to be the result of calibrator value miscalculation. Calibrator lot 3JD018 has now been value assigned to align with the correct calibrator values.

This poses a risk of an incorrect value of -10% being assigned for Phenytoin testing and 3% for Phenobarbital.

Siemens considers these differences are negligible and would not be considered substantially relevant to impact clinical care in practice.
Recall Action Recall for Product Correction
Recall Action Instructions Calibrator value reassignments are being provided in the Customer Letter for users to correct the problem. This action has been closed-out on 14/02/2017.