Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00061-1
Product Name/Description Medtronic Minimed 640G Infusion Pump used in clinical trials

Model Numbers: MMT-1511 & MMT-1512
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 29/01/2015
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue Medtronic has implemented an amendment to the Pump User Guide for Medtronic MiniMed 640G Insulin Pumps. The change involves the bolus delivery accuracy listed in the Pump User Guide for very small bolus amounts (less than 0.1 U). Delivery accuracy for bolus volumes < 0.1 unit is now updated to ±20%, from ±5%. The delivery accuracy for bolus volumes = 0.1 unit remains ±5%, as does the delivery accuracy for basal delivery.

Medtronic considers that there is not likely to be any clinical risk associated with this change in delivery accuracy.
Recall Action Recall for Product Correction
Recall Action Instructions An update to the published Instructions for Use (IFU) regarding the stated resolution of the device at a dose of <0.1 Unit is to be circulated to customers. This action has been closed-out on 08/08/2016.
Contact Information 02 9857 9000 - Medtronic Regulatory Affairs