Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00036-1
Product Name/Description Xcela PICC with PASV Valve Technology/Kits, BioFlo PICC with Endexo and PASV Valve Technology/Kits

Multiple lot numbers

ARTG numbers: 175080, 178780 & 203831
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/01/2015
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue Navilyst Medical is recalling PICC catheters that contain valve(s) manufactured prior to October 9, 2013. Use of these PICC catheters may have an increased incidence of hemolyzed blood samples which may require additional peripheral access or replacement of the PICC catheter.

Valves manufactured after this date incorporate a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter and they are not subject to this recall action.

Types of tests affected include Blood Chemistries such as electrolytes and Blood count measurements such as hemoglobin and hematocrit as they can be altered by hemolysis.

To date, no patient injuries have been reported to Navilyst Medical as a result of this issue.
Recall Action Recall
Recall Action Instructions Customers are asked to inspect and quarantine affected stock for return to Navilyst Medical as per the directions in the Customer Letter. This action has been closed-out on 18/07/2016.