Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00033-1 |
Product Name/Description |
Hand-held Barcode Scanner used with VITROS clinical chemistry systems. An in vitro diagnostic medical device (IVD).
Product Code: 6844210
ARTG Number: 180199 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
14/01/2015 |
Responsible Entity |
|
Reason/Issue |
Ortho Clinical Diagnostics (OCD) has received a report that when using a Hand-held Barcode Scanner in combination with use of the Batch Programming option for programming samples, the scanner unexpectedly skips sample cup positions. For example, a sample tube barcode is scanned for the Cup 1 position, the cursor in the Batch programming screen skips the Cup 2 position and advances to the Cup 3 position. If this anomaly occurs, there is a potential for mis-associated results. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
OCD is providing instructions on the reprogramming of the barcode scanner. Once the barcode scanner is reprogrammed the issue will not occur. This action has been closed-out on 15/07/2016 |
Contact Information |
1800 032 359 - Ortho Clinical Diagnostics Customer Technical Services |