Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00013-1 |
Product Name/Description |
Persona (TASP) Tibial Articular Surface Provisional Shim
All sizes and lots
Catalogue Number: 42-5279-003-00 thru 42-5279-009-04
ARTG number: 109936 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
15/01/2015 |
Responsible Entity |
|
Reason/Issue |
Zimmer is initiating this Recall due to Provisional Shims being found with missing ball bearings. Zimmer’s investigation has found that repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim.
This gives rise to the potential for delays in surgery and/or a missing part from the device being left in the wound.
This issue was also identified following worn returns being reported through the worn instrument program.
No complaints from Australia have been received regarding this issue. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Customers are requested to inspect any product referenced within this notification before and after use for missing ball bearings and to provide any devices with missing ball bearings to your Zimmer sales representative for return. This action has been closed-out on 18/03/2016. |
Contact Information |
02 9483 5400 - Zimmer Pty Ltd |