Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01358-1
Product Name/Description MEVATRON, PRIMUS, ONCOR and ARTISTE Digital Linear Accelerators

Multiple product codes

All serial numbers

ARTG number: 165502
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/12/2014
Responsible Entity Siemens Ltd
Reason/Issue Siemens has received reports of one circuit board (G41) located inside of the linac gantry showing signs of overheating. After a detailed analysis the root cause was found to be one of the components that belong to this board (Head Driver PCB). Siemens has redesigned the Head Driver PCB to further mitigate the risk of local overheating on the G41 motherboard.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens has redesigned the Head Driver PCB to further mitigate the risk of local overheating on the G41 motherboard. The present corrective action comprises the replacement of the Head Driver PCB by a redesigned component on your LINAC. This action has been closed-out on 12/08/2016.