Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01303-1
Product Name/Description Artis Zeego Systems (Fluoroscopic angiographic x-ray system)

C-arm Serial Number Range: 1000 to 1599

ARTG Number: 102177
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/12/2014
Responsible Entity Siemens Ltd
Reason/Issue A cable inside the c-arm of systems affected may have been routed in an inappropriate manner and there is a potential that with increased wear a cable inside the c-arm may break. A cable break will result in restricted functionality and even system failure.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is replacing the affected cable harness with a modified version in addition to re-routing the cables to prevent wear. This action has been closed-out on 01/03/2016.