Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01240-1
Product Name/Description LIAISON Borrelia IgG. An in vitro diagnostic medical device (IVD).

Part number: 310880

Lot Number: 089054X , 089054X/1 and 089054X/2

Expiry date January 9th, 2016 for all three lots

ARTG number: 208811
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 21/11/2014
Responsible Entity DiaSorin Australia Pty Ltd
Reason/Issue DiaSorin has become aware that the kit lots identified have a potential to produce invalid calibrations due to either a Calibrator 1 Deviation percentage [%] outside the high limit, or due to the GCC (Geometrical Curve Check) out of range. As a consequence the calibration cannot be validated and the testing with these lots on the analyser cannot be performed.
Recall Action Recall
Recall Action Instructions Diasorin is recommending that use of the affected lot be ceased and any units be discarded. Results obtained with a valid calibration are considered to be reliable; therefore, review of past patient results is not deemed necessary, but at the discretion of the laboratory director. This action has been closed-out on 01/03/2016.
Contact Information 1300 259 835 - SiaSorin Australia Pty Ltd