Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-01240-1 |
Product Name/Description |
LIAISON Borrelia IgG. An in vitro diagnostic medical device (IVD).
Part number: 310880
Lot Number: 089054X , 089054X/1 and 089054X/2
Expiry date January 9th, 2016 for all three lots
ARTG number: 208811 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
21/11/2014 |
Responsible Entity |
|
Reason/Issue |
DiaSorin has become aware that the kit lots identified have a potential to produce invalid calibrations due to either a Calibrator 1 Deviation percentage [%] outside the high limit, or due to the GCC (Geometrical Curve Check) out of range. As a consequence the calibration cannot be validated and the testing with these lots on the analyser cannot be performed. |
Recall Action |
Recall |
Recall Action Instructions |
Diasorin is recommending that use of the affected lot be ceased and any units be discarded. Results obtained with a valid calibration are considered to be reliable; therefore, review of past patient results is not deemed necessary, but at the discretion of the laboratory director. This action has been closed-out on 01/03/2016. |
Contact Information |
1300 259 835 - SiaSorin Australia Pty Ltd |