Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01205-1
Product Name/Description Stellaris PC with Laser System (Cataract extraction/vitrectomy system)

Product Code: BL14304
Serial Number : SPC02059

ARTG Number: 127264
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/11/2014
Responsible Entity Bausch & Lomb Australia Pty Ltd
Reason/Issue The Stellaris PC has a software fault, when in surgery, if the surgeon requests a Laser sub-mode change (i.e. from Endo to Endo Continuous) and the system is in “Standby”, the system will automatically transition from “Standby” to “Ready” to fire. If the system moves to Ready from Standby, and this has not been specifically commanded by the Surgeon or the Nurse, it raises the possibility that this change may not have been noticed and that the foot pedal could be depressed and the laser may fire unexpectedly.
Recall Action Recall for Product Correction
Recall Action Instructions Bausch and Lomb has developed a software update to address the situation, which will be installed in the affected system. In the meantime, users may continue to use the Stellaris PC laser system. However, they are advised not to request the Laser sub-mode change from Endo to Endo Continuous during surgery until the software update has been implemented. This action has been closed-out on 06/06/2016.