Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01192-1
Product Name/Description Dimension Vista 500 Intelligent Lab System & Dimension Vista 1500 Intelligent Lab System

Dimension Vista 500 Intelligent Lab System
Catalogue Number: 10488224

Dimension Vista 1500 Intelligent Lab System
Catalogue Number: 10444802

ARTG Number: 174699
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/11/2014
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens has received customer complaints of discrepant flagged and un-flagged patient and QC results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare Diagnostics has confirmed that reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing additional instructions to assist laboratories in determining when the replace the reagent probe.
Contact Information 1800 310 300 - Siemens Technical Support Center