Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01151-1
Product Name/Description Rocket IPC Drainage Bottle (A collection device used to collect fluid or air from the pleural cavity)

Product Codes: R54400, R54400-00-PP, R54410, R51400, R51400-00-PP, R51410

ARTG Number: 145658
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/11/2014
Responsible Entity Rocket Medical Pty Ltd
Reason/Issue Rocket Medical is notifying customers the blue connector tube may not open once the white clip is removed due to a manufacturing irregularity. If this occurs the vacuum contained in the bottle will not be applied to the patient, therefore no fluid can be drained. If left, the connector tube will still open with a delay of up to 5-6 minutes.
Recall Action Recall for Product Correction
Recall Action Instructions Rocket Medical is advising customers to open the tube by gently manipulating the tube forwards and backwards when the white clip has been opened. This action has been closed-out on 01/03/2016.