Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01112-1
Product Name/Description Artis zeego systems with a C - arm (Fluoroscopic X-ray systems)

C - arm serial number range: 1000 to 1599

ARTG number: 102177
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/10/2014
Responsible Entity Siemens Ltd
Reason/Issue Siemens have identified that a cable inside the C - arm of affected systems may have been routed in an inappropriate manner and there is a potential that with increase wear the cable inside the C - arm may break. A cable break will result in restricted functionality or system failure.
Recall Action Recall for Product Correction
Recall Action Instructions A Siemens service technician will contact affected customers to organise the modification of the cable routing to prevent the issue from occurring. This action has been closed-out on 14/07/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre