Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01079-1
Product Name/Description Selector Tubing Kit for use with CUSA Selector Ultrasonic Aspirators

Catalogue number: 1517079

Multiple lot numbers

ARTG number: 146847
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 9/10/2014
Responsible Entity Integra Neurosciences Pty Ltd
Reason/Issue Integra NeuroSciences has identified through an investigation of complaints that some specific lots of SELECTOR® Tubing Kits may potentially leak irrigation fluid during use if not placed on the SELECTOR handpiece in a specific manner. If the lumens of the 2 tubes are not separated before placing them on the handpiece barbs, this could result in an irrigation fluid leak. If the leakage is significant, the handpiece tip could overheat and possibly burn tissue being treated.
Recall Action Recall for Product Correction
Recall Action Instructions Integra is providing additional instructions to prevent the occurrence of leakage. This action has been closed-out on 03/06/2016.