Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-01078-1 |
Product Name/Description |
IMRIS Operating Room table (ORT300)
Model number: 114093-000
Lot numbers: 10002255
ARTG number: 206814 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
7/10/2014 |
Responsible Entity |
|
Reason/Issue |
Emergo Australia is replacing the table alignment block pins, which are currently made from ferrous materials, with pins made from non-ferrous material due to an incident at a health care facility in the United States. One of the three pins that is used to hold the ORT200 table alignment block in place on the floor of the MR Suite went missing and was not accounted for by OR personnel. During testing the missing pin was pulled into the magnet, damaging the image quality phantom and headrest extension. No injury or death occurred as a result of this incident. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Emergo Australia is replacing the table alignment block pins, which are currently made from ferrous materials, with pins made from non-ferrous material. A field engineer will be sent to the customer site to replace the pins. This action has been closed-out on 06/06/2016. |
Contact Information |
02 9006 1662 - Emergo Australia |