Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01065-1
Product Name/Description Autostainer Link 48, AS480

Software version: Dako Link software 4.0.3
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/10/2014
Responsible Entity Dako Australia Pty Ltd
Reason/Issue Dako has found a software defect in the calibration routine of DakoLink 4.0. The defect may cause the Autostainer control software to use old calibration settings.
Recall Action Recall for Product Correction
Recall Action Instructions Dako implemented a software update to correct the issue. This action has been closed -out on 28/01/2016

This action was undertaken prior to consultation with the Therapeutic Goods Administration.
Contact Information 02 9922 0700 - Dako Australia