Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01030-1
Product Name/Description Dimension Vista Hemoglobin A1c Flex Reagent Cartridge. An in vitro diagnostic medical device (IVD)

Catalogue Number: DF105A

Siemens Material Number: 10483822

Lot Numbers: GA4266, BA4273, BA4280, BA4287, GA4301, GA4315, GC4322, GA4343, GA4350, GA4357, GA5013 & GA5020

ARTG Number: 181689
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/09/2014
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens has confirmed that the Dimension HB1C Flex reagent cartridge lots listed1 exhibit a positive bias averaging 0.4% [4.4 mmol/mol] Haemoglobin A1c units and occasionally up to 1.0% [11 mmol/mol] HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP).
Recall Action Recall
Recall Action Instructions Siemens is advising their customers to discontinue use and discard the affected lots. Retesting HbA1c should be considered in cases where all of the following events have occurred:
- HbA1c testing was performed on your patient(s) using the affected lot numbers
- There may have been adjustments in therapy based solely on the HbA1c value(s),
- There have been no follow up HbA1c testing on the patient since the discontinuation of use of the affected lot numbers. This action has been closed-out on 10/08/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre