Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01003-1
Product Name/Description Aeroneb Pro/Solo Nebuliser Systems FRIWO power adapter

Multiple catalogue numbers and date codes affected

ARTG Number: 99080
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/09/2014
Responsible Entity Tag Medical Pty Ltd
Reason/Issue Tag Medical have been advised by Aerogen Ltd that there is a potential hazard with certain FRIWO power adaptors as supplied with the Aerogen Pro and Aerogen Solo nebuliser systems manufactured during the period of week 50/2013 and week 23/2014. The power adapter could potentially fail, whereby the plastic screw pillars can break and dislodge from the body and the adapter cover can open.
Recall Action Recall for Product Correction
Recall Action Instructions Tag Medical is contacting the hospitals and replacing the faulty FRIWO power adaptors free of charge. This action has been closed-out on 06/06/2016.