Type of Product Medical Device
TGA Recall Reference RC-2014-RN-01001-1
Product Name/Description VITROS 4600 Chemistry Systems & VITROS 5600 Integrated Systems Using Software Version 3.0

VITROS MicroSlide Calibrator Kit Numbers: 1 - 10 (inc.) & 25

ARTG Number: 180199
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/09/2014
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue An unexpected assay calibration may occur when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits for assay targets that are unassigned.
Recall Action Recall for Product Correction
Recall Action Instructions A correction for this issue will be included in the next version of the software. Ortho-Clinical Diagnostics is requesting their customers to not use the calibrator barcode labels for VITROS Calibrator Kits 1, 2, 3, 4, 6, or 9 until the next version of software is installed on the VITROS System. This action has been closed-out on 25/08/2016.
Contact Information 1800 032 359 - Ortho-Clinical Diagnostics Customer Technical Services