Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00994-1
Product Name/Description 942-073, D8088_Crea A, D8089_Crea B Membranes used on the ABL827 blood gas analyser. An in vitro diagnostic medical device (IVD)

Lot Number: R0530
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 15/09/2014
Responsible Entity Radiometer Pacific Pty Ltd
Reason/Issue Radiometer Medical has become aware that Lot R0530 of 942-073, D8088-Crea A,D8089-Crea B, membrane box for E8088 and E8089 electrodes has been packed in error with capsules of electrolyte solutions for Glucose/Lactate electrodes. Testing by the manufacturer has established that the incorrect capsules of electrolyte solutions do not influence the performance of the Crea results.
Recall Action Recall
Recall Action Instructions Radiometer is requesting their customers to discard remaining units from the affected lots. Alternative units are available. This action has been closed-out on 01/03/2016.
Contact Information 1800 247 254 - Radiometer Pacific