Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00971-1
Product Name/Description Salfo Talaris Tibial Impactor

Part Number: MJU361 (This instrument is also part of the Salto Talaris Instrumentation, Part Number YKAL13)

Serial Numbers: 30, 152

ARTG Number: 112287
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 18/09/2014
Responsible Entity Tornier Pty Ltd
Reason/Issue Complaints have been received which describe a non-optimum grip of tibial inserts with the tibial impactor Salto Talaris which could lead to the fall of the implant. Tornier's investigations have led to the conclusion that only sizes 0 and 1 of the tibial implant are not sufficiently maintained by the tibial impactor which, however, remains functional.
Recall Action Recall for Product Correction
Recall Action Instructions Tornier advised their customers that even if the impactor does not ensure optimal holding of sizes 0 and 1 only of the tibial implant, it can still safely be usedand requested that care be taken when using the device for implantation of sizes 0 and 1 of the tibial implants. Tornier is redesigning the device and will notify customers once complete.