Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00963-1 |
Product Name/Description |
Spectra Optia Apheresis System with Software Version 11
Catalogue Number: 61000
ARTG Numbers: 130529 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
5/09/2014 |
Responsible Entity |
|
Reason/Issue |
A defect in Optia version 11 SW can cause an unexpected behaviour to the rinseback option when the WBC or platelet counts are updated during the run and they are sufficiently different from the previous counts. The absence of rinseback could result in loss of RBC potentially resulting in limited illness or injury. Unintended rinseback carried out on a very small patient could result in symptomatic hypervolemia. No serious injury or deaths have been associated with this behaviour. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Terumo BCT is actively working on a software update to correct this behaviour. Once this software is available affected devices will be updated. End users are being provided work around instructions to ensure the continued safe use of the device. This action has been closed-out on 14/07/2016. |
Contact Information |
02 9429 3600 - Terumo BCT Australia |