Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00961-1
Product Name/Description Babytherm 8004/8010 and Babytherm 8000WB for newborns (Infant warmers and open infant care units)

ARTG Number: 121161
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 9/09/2014
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue Dräger have become aware of a case in which a side panel on a Babytherm 8010 did not fully slide to the bottom in the corner guide and was not completely locked. The side panel did not open during this incident and no patient was affected. As part of the review, Dräger identified that the inner 70mm panels were not always used however, the inner 70mm panels are part of the safety concept of the Babytherm 80XX. If contrary to the information provided in the Instructions for Use Manual the inner panels are not inserted and if a side panel has not been fully locked into position this may allow the baby’s movements to open the side panel. It may be possible for a child to fall out.
Recall Action Recall for Product Correction
Recall Action Instructions Users are reminded to pay attention to the information provided in the Instruction for Use Manual: i.e., “Check that the side panels lock securely into position and that they remain securely locked”. If the locking mechanism is not functioning or the inner panels are missing or damaged, please contact Dräger Customer Service to arrange correction. This action has been closed-out on03/02/2017.
Contact Information 03 9244 7248 - Drager Medical