TGA Recall Reference RC-2014-RN-00916-1
Product Name/Description Vapotherm Disposable Patient Circuit System (Delivers high flow humidified gases via a nasal cannula or other patient interface)

Part Numbers: PF-DPC-High, PF-DPC-Low

Lot Numbers: HF1402035, HF1404043, LF1402030, LF1405023

ARTG Number: 147794
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/08/2014
Responsible Entity Surgical Specialties Pty Ltd
Reason/Issue Vapotherm Inc. has received a small number of complaints involving a defect in the Disposable Patient Circuit Delivery Tube. The defect allows water to leak into the centre lumen of the delivery tube, the gas line to the patient. The fault condition is easily detectable and typically occurs while the device is coming up to the set temperature and is easily visible when following the Precision Flow instructions for use, which requires that the cannula be connected to the patient only after the set temperature is reached.
Recall Action Recall
Recall Action Instructions Customers are asked to check their stock immediately to determine if they have any affected product. Customers are advised to return the affected stock to Surgical Specialties Pty Ltd. This action has been closed-out on 10/05/2017.